OT-DICE Manual


6.2 Sample Session 2

Test Scenario

Checks the Echo SCU and Echo SCP services. In this case the Storage SCP service takes the role of the Echo SCP service.

Steps

  1. Keep the configuration details created for Sample Session 1 for both the Storage SCP node and the DICOM node (test_node).

  2. Start Storage SCP mode, if it is not already started.

  3. Switch to Echo SCU mode by either clicking on the ECHO SCU button in the Mode window of the main window or by selecting Echo SCU from the Modes menu.

  4. The storescp value will appear in the Called AE Title combo box in the main window.

  5. Select the storescp value in the Called AE Title combo box in the main window.

  6. Finally, click on the SEND button to initiate an Echo request.

Results

The echo succeeds and the log window displays the status of the operations. 

The Echo SCU Local Result:

 

The Echo SCU Remote Result:

 

The Log:

 

Return to OT-DICE User Manual Table of Contents

OTech, Inc., OT-DICE User Manual